Electrical equipment for medicine
|
GROUNDING SOCKET
|
POST OF REMOTE CONTROL
|
Modern medical technologies lean on numerous systems of diagnostics, treatment and rehabilitation of the patient. At use of modern electronic equipment in medical process it is important not only reliability and accuracy of operation of the equipment, but also its safe application.
At a traditional way of power supply of medical equipment (the grounded neutral) electrical safety can be not provided. Damage of isolation of electric equipment can lead to defeat by electric current of the patient and medical personnel. Besides, shutdown of the line of power supply to which the systems providing activity of the patient can be connected is possible.
The issue of electrical safety is resolved by means of the correct catering services of medical electric equipment – the organization of a network with the isolated neutral (IT network). For transformation of a TN-S network to an IT network medical dividing transformers are used.
Control of a condition of a network and the medical dividing transformer carries out the panel of control of operational PKO-03, and the USP FK uninterruptible power supply carries out power supply of surgical lamps.
For connection of the medical equipment rossete guards of various complete set, both with single-phase, and with three-phase sockets, necessary number of machine guns, grounding plugs, indication of tension are used. We offer hinged plastic guards of ESR and the built-in metal guards of SRM.